New Olympus EPOCH 6LT Portable Flaw Detector

New Olympus EPOCH 6LT Portable Flaw Detector

General

User Interface LanguagesEnglish, Spanish, French, German, Japanese, Chinese, Portuguese, Russian, Italian
Transducer ConnectionsLEMO 00
Data Storage100,000 IDs onboard
Battery Type and LifeSingle lithium-ion rechargeable standard; 6 h life
Power RequirementsAC Mains: 100 VAC to 120 VAC, 200 VAC to 240 VAC, 50 Hz to 60 Hz
Display TypeFull VGA (640 × 480 pixels) transflective color LCD, 60 Hz update rate
Display Dimensions (W × H, Diag.)117 mm × 89 mm, 146 mm (4.62 in. × 3.49 in., 5.76 in.)
Overall Dimensions (W × H × D)209 mm x 128 mm x 36 mm, 58 mm at hand grip (8.2 in. x 5 in. x 1.4 in., 2.3 in. at hand grip)
Weight890 g (1.95 lb), including lithium-ion battery

Instrument Inputs/Outputs

USB Ports(1) USB 1.1 Full Speed Host (Type A)
(1) USB 2.0 Full Speed Client (Type Mini B)
Video Output1 digital video output

Environmental Ratings

IP RatingIngress Protection (IP) engineered to IP67 (dust tight and water submersion) and IP65 (dust tight and water jets) per IEC 60529-2004 (Degrees of Protection provided by enclosures—IP Code). The product rating is confirmed by means of Olympus’ internal design verification test process.
Explosive AtmosphereMIL-STD-810F, Method 511.4, Procedure 1.
Shock TestedMIL-STD-810F, Method 516.5, Procedure I, 6 cycles each axis, 15 g, 11 ms half sine.
Vibration TestedMIL-STD-810F, Method 514.5, Procedure I, Annex C, Figure 6, general exposure: 1 hour each axis.
Operating Temperature-10 °C to 50 °C (14 °F to 122 °F)
Battery Storage Temperature0 °C to 50 °C (32 °F to 122 °F)

Pulser

PulserTurnable square wave
PRF10 Hz to 2000 Hz in 10 Hz increments
Energy Settings100 V, 200 V, 300 V, or 400 V
Pulse WidthAdjustable from 25 nsec to 5,000 nsec (0.1 MHz) with PerfectSquare technology
Damping50, 400 Ω

Receiver

Maximum Input Signal20 Vp
Receiver Input Impedance400 Ω ± 5%
Receiver BandwidthDC to 26.5 MHz at -3 dB (standard version)
0.2 to 26.5 MHz at -3 dB (ISO 22232-1:2020 compliant version)
Digital Filter Settings8 digital filter sets (standard version)
7 digital filter sets (ISO 22232-1:2020 compliant version)
RectificationFull-Wave, Positive Half-Wave, Negative Half-Wave, RF
System LinearityHorizontal: ± 0.5% FSW
Resolution0.25% FSH, Amplifier Accuracy /- 1dB
Reject0 to 85% FSH in 1% increment positions
Amplitude Measurement1.25% to 110% full screen height
Measurement RateEquivalent to PRF in all modes (single shot)

Calibration

Automated CalibrationVelocity, zero offset
Straight beam (first back wall or echo-to-echo)
Angle beam (sound path or depth)
Test ModesPulse Echo, Dual, or Through Transmission
UnitsMillimeters, inches, or microseconds
Range4.31 mm to 6,700 mm at 5,900 m/s (0.2320 in./μs)
Velocity635 m/s to 15240 m/s (0.0250 in./μs to 0.6000 in./μs)
Zero Offset0 to 750 μs
Display Delay-10 microseconds to 2203 microseconds
Refracted Angle0° to 85° in 0.1° increments, then jump to 90°

Gates

Measurement Gates2 fully independent flaw gates
Gate StartVariable over entire displayed range
Gate WidthVariable over entire displayed range
Gate HeightVariable from 2 to 95% full screen height in 1% increments
AlarmsPositive and negative threshold/curve, minimum depth (gate 1 and gate 2)

Measurements

Measurement Display Locations5 locations available (manual or auto selection)
GATE (1, 2)Thickness, sound path, projection, depth, amplitude, time-of-flight, min./max. depth, min./max. amplitude, sizing measurements based on mode
Measurement Display LocationsUser selects up to four measurements from either gate to display on the live screen.
Echo-to-echoStandard gate 2 – gate 1
DAC/TCGStandard, up to 50 points, 110 dB dynamic TCG range
Special DAC ModesCustom DAC (up to 6 curves), 20–80% view
Curved Surface CorrectionStandard OD or bar correction for angle beam measurements
admin Avatar

Leave a Reply

Your email address will not be published. Required fields are marked *