New Olympus EPOCH 650 Ultrasonic Flaw Detector

New Olympus EPOCH 650 Ultrasonic Flaw Detector

USB On-The-Go (OTG)
Yes
VGA output standard
1 analog output (optional), Selectable 1 V/10 V Full Scale, 4mA max
3 alarm outputs, 5 V TTL, 10 mA
Trigger input, 5V TTL;
Trigger output, 5V TTL, 10 mA max
1-axis encoder line (quadrature – Corrosion Module mode only)Designed to meet the standards of the Ingress Protection (IP) rating number IP67 (navigation pad version) or IP66 (knob version) per IEC 60529-2004 (Degrees of Protection provided by enclosures – IP Code).The product design was confirmed to meet the IP rating by means of Olympus internal design verification test process that occurs prior to the release of the product to production.

Safe operation as defined by Class I, Division 2, Group D, as defined in the National Fire Protection Association Code (NFPA 70), Article 500, and tested using MIL-STD-810F, Method 511.4, Procedure I.
MIL-STD-810F, Method 516.5 Procedure I, 6 cycles each axis, 15g, 11 ms half sine.
MIL-STD-810F, Method 514.5, Procedure I, Annex C, Figure 6, general exposure: 1 hour each axis
–10 °C to 50 °C (14 °F to 122 °F)
0 °C to 50 °C (32 °F to 122 °F)236 mm x 167 mm x 70 mm (9.3 in. x 6.57 in. x 2.76 in.)
1.6 kg (3.5 lb), including lithium-ion battery
English, International, Japanese, Chinese
English, Spanish, French, German, Japanese, Chinese, Portuguese, Russian
BNC or Number 1 LEMO
100,000 IDs onboard, removable 2 GB MicroSD card (standard)
Single lithium-ion rechargeable standard
15 h to 16 h (lithium-ion)
AC Mains: 100 VAC to 120 VAC, 200 VAC to 240 VAC, 50 HZ to 60 Hz
Full VGA (640 x 480 pixels) transflective color LCD, 60 Hz update rate
117 mm x 89 mm, 146 mm (4.62 in. x 3.49 in., 5.76 in.)Tunable Square Wave
10 Hz to 2000 Hz in 10 Hz increments
100 V, 200 V, 300 V or 400 V
Adjustable from 25 ns to 5,000 ns (0.1 MHz) with PerfectSquare™ technology
50, 100, 200, 400 Ω0 to 110 dB
20 V p-p
400 Ω ± 5%
0.2 MHz to 26.5 MHz at -3 dB
Thirty digital filter sets standard
Seven ISO 22232-1:2020 compliant filters (0.2-10 MHz, 2.0-21.5 MHz, 8.0-26.5 MHz, 0.5-4 MHz, 0.2-1.2 MHz, 1.5-8.5 MHz, 5-15 MHz)
Full-wave, Positive Half-wave, Negative Half-wave, RF
Horizontal: ± 0.5% FSW
0.25% FSH, amplifier accuracy ± 1dB
0 to 80% FSH with Visual Warning
0 to 110% full screen height with 0.25% resolution
Equivalent to PRF in all modesVelocity, Zero Offset
Straight Beam (First Backwall or Echo-to-Echo)
Angle Beam (Soundpath or Depth)
Pulse Echo, Dual, or Through Transmission
Millimeters, inches, or microseconds
3.36 mm to 13,388 mm (0.132 in. to 527.10 in.) @ 5,900 m/s (0.2320 in./μ)
635 m/s to 15240 m/s (0.0250 in./µs to 0.6000 in./µs)
0 µs to 750 µs
-59 mm to 13,401 mm (-2.320 in. to 526.97 in.) @ longitudinal velocity in steel
0° to 90° in 0.1° increments2 fully independent gates for amplitude and TOF measurements
Variable over entire displayed range
Variable from Gate Start to end of displayed range
Variable from 2 to 95% full screen height
Positive and Negative Threshold, Minimum Depth (Gate 1 and Gate 2)5 locations available (manual or auto selection)
Thickness, Soundpath, Projection, Depth, Amplitude, Time-Of-Flight, Min./Max. Depth, Min./Max. Amplitude
Standard Gate 2-Gate 1, Optional IF Gate Tracking
Overshoot (dB) value for DGS/AVG, ERS (equivalent reflector size) for DGS/AVG, AWS D1.1/D1.5 A, B, C and D values, Reject Value, Echo to RefdB values
Standard
Up to 50 points, 110 dB dynamic range
Custom DAC (up to 6 curves), 20-80% View
Standard OD or Bar correction for Angle Beam measurements
Zero-cross measurement algorithm, V-Path correction, Single or Echo-to-Echo, Encoded B-scan

admin Avatar

Leave a Reply

Your email address will not be published. Required fields are marked *